NedPak

Technological Solutions For Post–Harvest and Packaging

If you don't measure Ethylene, you can't control it!

Gas Analyser - O2/CO2 Analyzer

Map-Pak Analyser Combi Series

Accurate Analysis of Gas Flushed Packaging


Features
  • Large LCD Display with backlight
  • Measurement range of 0.1% - 100% with accuracy of ± 1%
  • Measurement results achieved withinseconds
  • Integrated data logging of up to 5,000 measurements
  • Adjustable alarm feature with pass/fail indicator
  • Safety holster for needle/probe
  • Operator I.D. available
  • Product I.D. selectable
  • Availablein multiplelanguages

About the new Map-Pak Combi

Designed specifically for the Food Industry, the Map-Pak Combi measures the Oxygen, Carbon Dioxide and Nitrogen content in gas flushed food packages (the latter being as a balance gas) It can be used for all food applications, as well as some applications in the pharmaceutical industry.

This lightweight portable analyser has a range of features, making it easier to use in quality control functions. Sample measurements and results are achieved within 10 seconds and the operator can modify the various features of the analysers to best suit your requirements.

The Map-Pak Combi is time and cost efficient. It offers all round quality with top of the range features as well as functional features which are designed to make the Map-Pak Combi user friendly.

What does the Map-Pak Combi Offer ?

  • Lightweight & Portable
  • Fast and Accurate O2 & CO2 &N2 readings
  • Easy to useinterface
  • Minimal gassample required
  • Easy tocalibrate
  • 7 hours ofBattery Life/Alternativemains power supply
  • Fast recharge feature
  • Needle blockage alarm

Principle Of Operation

The Map-Pak Combi, using a small gas sample from the gas flushed pack, measures and records levels of Oxygen and Carbon Dioxide. The levels of gas are measured using specially customised sensors; Oxygen levels are measured by an electrochemical sensor and the Carbon Dioxide levels are measured by a nondispersive infrared sensor (NDIR). An electric pump provides a small amount of gas into the sensor. The O2 and CO2 particles are detected by the respective sensors and the measurements are displayed on the screen within seconds. The N2 level is then displayed as a balance gas reading.

Specification
<
AccuracyO2 ± 1% (full scale) CO2 ± 5%
Resolution0.1%
Repeatability1% relative to gas reading
Measuring MethodElectro Chemical Sensor (O2), Nondispersive infared Sensor(C2)
Measuring Range 0% to 100% in 0.1% increments
Ambient Temperature0 to +40 degrees C
Response Time <10 seconds
Sensor LifetimeO2 Sensor -2 years CO2 Sensor-Inde?nitely
Minimum Sample Gas Requirement5 ml
Sample Probe Probe with ?exible sampling connection to internal pump
Alarms Alarms feature thresholds: can be de?ned seperately for every product by the customer
Calibration Simple 2 point Calibration ( O2 Sensor Only )
User Interface Graphic Display with backlight 30 x 60 [mm]
Data Output Serial Interface-USB
Data StorageUp to 5,000 measurements can be stored internally
Power Supply
Mains Adapter9V wall adapter (Universal AC supply)
Battery Pack 4 NiMH rechargeable cells
Operation Time 7 hours normal use from full charge / Continuous operation using mains adapter
Power SaverAutomatic Shutdown after 2 minutes of non-use/idle time
EnclosureABS Plastic
Weight 450 grams
DimensionsL =228, W=72/117, H=47 [mm]
Approvals CE - marked according to:
- EMC 2004/108/ EC - Low Voltage Directive 2006/95/EC

Aftersales Care

AGC Instruments are committed to providing and maintaining a quality product from the point of sale. We pride ourselves on our dedicated after sales service in order to give our customers value for their purchase.

Additional Accessories Included

The Map-Pak Combi comes complete with Needles (5 off), Septums (100 off), Particle Filters (2 off), Needle Probe (1 off), Carry Case (1 off), User Manual (1 off), Rechargeable Batteries (4 off) and Mains Adapter (1 off).